Medical Device Safety Alert: Medtronic O-arm 1000 Imaging Systems

10 Sep 2018

Medical device manufacturer, Medtronic, has issued a medical device safety alert concerning all 2nd Edition O-arm 1000 Imaging System. The affected devices are identified as follow:

  • Product Names (Catalogue Number)
    • OARM ASSY BI70000027 SYSTEM PRODUCT (BI70000027)
    • OARM ASSY BI70000027R SYS PRODUCT RWK (BI70000027R)
    • BASE OARM BI70000027100 SYSTEM 100V (BI70000027100)
    • BASE OARM BI70000027100R SYSTEM 100V RWK (BI70000027100R)
    • BASE OARM BI70000027120 SYSTEM 120V (BI70000027120)
    • BASE OARM BI70000027120R SYSTEM 120V RWK (BI70000027120R)
    • BASE OARM BI70000027230 SYSTEM 230V (BI70000027230)
    • BASE OARM BI70000027230R SYSTEM 230V RWK (BI70000027230R)
    • OARM ASSY BI70000027GER SYS PRODUCT GER (BI70000027GER)
    • OARM ASSY BI70000027GERR SYS PRODUCT REFURB (BI70000027GERR)

The manufacturer released software version 3.1.7. The change is in response to customer complaints and addresses several known software anomalies. The software anomalies were reported to affect the following system functionalities:

  • Startup and shutdown
  • System and network communication
  • Motion control
  • Image acquisition and output
  • Dose reporting
  • Logs generation

According to the manufacturer, over the last six years, five complaints led to cancelled procedures after the patient was under anaesthesia requiring additional surgery. The field action is being initiated to address the issues relating to these complaints.

The local supplier will contact affected users to schedule completion of the software install. The affected users may choose to continue to use their 2nd Edition O-arm 1000 Imaging System at their clinical discretion in the meantime, but they should be aware of the issues described above, which may result in a delayed surgery, patient exposure to non-navigated surgery, patient exposure to additional surgery or unused X-ray dose.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary actions.

Posted on 10 September 2018