Medical Device Safety Alert: Medtronic Covidien Signia Small Diameter Curved Tip Intelligent Reload

08 Mar 2021

Medical device manufacturer, Medtronic, has issued a medical device safety alert concerning its Covidien Signia Small Diameter Curved Tip Intelligent Reload. The affected devices are identified as the following: -

  • Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long
    • Item Code: SIGSDL45CTVT
    • Affected Lot Numbers: N0A1038UY N0E0016UY N0F0616UY N0K0934UY N0B0493UY N0F0254UY N0F0817UY N0M0334UY N0C0436UY N0F0518UY N0G0388UY N9K1095UY N0D0731UY N0F0615UY N0J0859UY N9L0797UY
  • Covidien Signia Grey Small Diameter Curved Tip Intelligent Reload30 mm Vascular 8 mm - Short
    • Item Code: SIGSDS30CTV
    • Affected Lot Numbers: N0A0191UY N0D0392UY N0F0406UY N0L0708UY N0B0657UY N0E0133UY N0G0907UY N9K0376UY N0C0349UY N0E0956UY N0J0746UY N9K1093UY N0C0492UY N0E1026UY N0K0248UY N9L0796UY N0C1008UY N0F0202UY N0K0803UY N9M0418UY N0D0283UY N0F0221UY N0L0307UY N9M0651UY
  • Covidien Signia Small Diameter Curved Tip Intelligent Reload 30 mm Vascular/Thin 8 mm - Short
    • Item Code: SIGSDS30CTVT
    • Affected Lot Numbers: N0A0274UY N0D0037UY N0G0491UY N0K1016UYN0A0897UY N0D0391UY N0G0736UY N0L0077UY N0B0640UY N0E0175UY N0J0510UY N0M0063UY N0B0937UY N0E0802UY N0J0680UY N0M0259UY N0C0144UY N0F1055UY N0J0999UY N9K1094UY N0C0486UY N0G0185UY N0K0584UY N9L0795UY N0C1115UY N0G0250UY N0K0805UY

Product recall is being conducted and is on-going due to the potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

The manufacturer is instructing affected users to immediately quarantine and discontinue use of affected devices.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary action.

Posted on 8 March 2021