Important Safety Alerts

Skip to main content

Safety Alerts and Communications

Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Beckman Coulter Access AMH Calibrators

21 Jun 2021

Medical device manufacturer, Beckman Coulter, has issued a medical device safety alert concerning its Access AMH Calibrators [Reference: B13128; Lot Number: 189303, 189304 of Expiry Date: 2022-07-31 and Lot Number: 189305, 189306 of Expiry Date: 2022-08-22].

The manufacturer has determined through an internal investigation that the Access AMH (i.e. Anti-Müllerian Hormone) Calibrator lots listed above do not meet the open vial stability claim when used on days 32-90 after reconstitution. According to the manufacturer, the open vial calibrator stability from 0 to 31 days is not impacted.

According to the manufacturer, the Access AMH calibrator instability could result in elevated quality control results of up to 20% across the reportable assay range. This could result in quality control failures, which could delay the reporting of results.

The manufacturer is advising affected users to take the following actions: -

  • Discontinue use of the Access AMH calibrator lots listed above beyond 31 days after reconstitution.
  • Review QC results if they generated a calibration curve from the affected calibrator lots that had been 32 days or longer since the reconstitution.

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary action.

Posted on 21 June 2021

Top
Level Double-A conformance, W3C WAI Web Content Accessibility Guidelines 2.1(Thumbnail) Brand Hong Kong(Thumbnail)