Important Safety Alerts

Skip to main content

Safety Alerts and Communications

Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Medtronic Symplicity Catheters

28 Jan 2013

Medical device manufacturer, Medtronic Ltd., has issued a field safety notice concerning Symplicity Catheters. The affected model numbers are IDERDN006 and RDN006.

The manufacturer has observed a connection issue complaint rate of approximately 5% of devices provided since enrollment of the Symplicity HTN3 trial began. These connection issues cause unstable temperature readings or lack of signal to the generator. In events where a connection failure leads to unstable temperature reading, the generator's safety algorithm is triggered and energy delivery is terminated, or if there is a lack of signal to the generator, it will not initiate treatment, thus in either situation protecting the patient from unsafe treatment.

These connection issues manifest in one of the following ways:

  • a flickering generator display screen upon initial catheter connection
  • generator status display of "check connection"
  • rapid fluctuations in the temperature reading
  • multiple early termination of energy delivery accompanied by "check status" safety codes 22, 23 and/or 32 despite confirming electrode placement as defined by the IFU.

The result of these connection failures is the potential need during the procedure to replace the catheter to successfully complete the renal denervation procedure.

The manufacturer advises users to connect the catheter into the generator prior to inserting into the patient according to the IFU. Furthermore, the manufacturer advises users that if observing any of the conditions above, they should follow the steps below:

  • Assure the catheter connector plug is fully seated into the generator by unplugging and re-plugging the catheter. If this resolves the issue, continue with treatment.
  • Gently wiggle the connector plug. If this resolves the issue, continue with treatment.
  • If the issue remains, replace the catheter and return the catheter to Medtronic for investigation.

Medtronic has also developed a new catheter connector.

According to the local supplier, Medtronic International Limited, the affected devices were distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary actions.

Posted on 28 January 2013

Top
Level Double-A conformance, W3C WAI Web Content Accessibility Guidelines 2.1(Thumbnail) Brand Hong Kong(Thumbnail)