Medical Device Safety Alert: Ortho Clinical Diagnostics VITROS Chemistry Products Specialty Diluent and VITROS Chemistry Products FS Diluent Pack 3

28 Mar 2014

Medical device manufacturer, Ortho Clinical Diagnostics, has issued a medical device safety alert concerning its VITROS Chemistry Products Specialty Diluent Lot F3168 (Product Code 8559825) & VITROS Chemistry Products FS Diluent Pack 3 Lot 01-3266 (Product Code 6801754).

The manufacturer confirmed reports of lower than expected C-Reactive Protein (CRP) results when using the affected lots of VITROS Specialty Diluent and VITROS FS Diluent Pack 3 to dilute samples tested on VITROS CRP Slides. When using the affected lots of diluent, results using VITROS CRP Slides may be negatively biased by approximately 51% using a 3X dilution factor.

The manufacturer advises affected users to take the following actions:-

According to the manufacturer, the affected products were not distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.