Medical Device Safety Alert: Ortho Clinical Diagnostics VITROS Immunodiagnostic Products NT-proBNP Calibrators, VITROS Immunodiagnostic Products NT-proBNP Reagent Packs

08 Apr 2018

Medical device manufacturer, Ortho Clinical Diagnostics, has issued a medical device safety alert concerning its VIITROS Immunodiagnostic Products NT-proBNP Calibrators (Product Code 6802157; Lot 1110) and VITROS Immunodiagnostic Products NT-proBNP Reagent Packs (Product Code 6802156; Lot 1110).

During routine internal testing, the manufacturer identified potential for non-reproducible, higher than expected results that may exceed the decision thresholds for congestive heart failure (CHF) when using VITROS NT-proBNP Reagent Packs, Lot 1110. When using the affected lots of products, elevated results were observed across the measuring range of the assay (5.00 to 35,000 pg/mL and 0.590 to 4130 pmol/L). Analysis determined that this issue may affect approximately 1 in 73 test results, or 1.37%.

The manufacturer advises affected users to take the following actions:-

According to the manufacturer, the affected products were not distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.