Medical Device Safety Alert: Philips Medical Systems Heartstart MRx Defibrillator/Monitor

05 Dec 2014

Health Canada posted a medical device safety alert concerning Heartstart MRx Defibrillator/Monitor, manufactured by Philips Medical Systems. According to the FSN from the manufacturer, the affected devices are HeartStart MRx Monitors/Defibrillators having EtCO2 Option and/or Old Cap Plate with the followings:

The manufacturer has become aware that the Heartstart MRx Defibrillator/Monitor can be susceptible to one or both of the issues described below:-

If the CO2 Inlet Port can no longer be accessed, the user will be unable to initiate EtCO2 monitoring when needed. Separation of the handle from the MRx housing can cause the MRx to fall, potentially injuring the patient or user. In addition, a delay of therapy could occur if the device is damaged by the impact.

The manufacturer will provide hardware upgrade for all affected units and will contact customers with affected devices to arrange for installation of the upgrade.

According to the local supplier, the affected products were distributed in Hong Kong.

For details, please refer to the following Health Canada website:
http://www.healthycanadians.gc.ca/recall-alert-rappel-avis/hc-sc/2014/42779r-eng.php

If you are in possession of the affected products, please contact your supplier for necessary actions.