Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: GE Medical Systems Magnetic Resonance Imaging Systems

08 Jan 2015

Health Canada posted a medical device safety alert concerning various Magnetic Resonance Imaging (MRI) systems, manufactured by GE Medical Systems. The affected devices are identified as follow:-

In the affected devices, Magnet Rundown Unit (MRU) may not be connected to the magnet. If the MRU is disconnected, the MRU will not function as expected and will not quench the magnet when the button is pressed. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.

For details, please refer to the following Health Canada websites:

http://www.healthycanadians.gc.ca/recall-alert-rappel-avis/hc-sc/2014/43153r-eng.php

http://www.healthycanadians.gc.ca/recall-alert-rappel-avis/hc-sc/2014/43157r-eng.php

http://www.healthycanadians.gc.ca/recall-alert-rappel-avis/hc-sc/2014/43103r-eng.php

http://www.healthycanadians.gc.ca/recall-alert-rappel-avis/hc-sc/2014/43165r-eng.php

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 8 January 2015

Back