Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Beckman Coulter Access Ostase Calibrators and Access Ostase QC Kits

30 Nov 2015

Medical device manufacturer, Beckman Coulter, has issued a medical device safety alert concerning its Access Ostase Calibrators and Access Ostase QC Kits. The affected products are identified as follows:

According to the manufacturer, the Access Ostase Calibrator and Access Ostase QC lots listed above do not meet their labeled expiration dates when stored after initial use (open vial stability). When stored unopened, calibrators and quality controls are stable until the expiration date on the vials. Through internal testing, the manufacturer has determined that the above lots are stable at 2°C to 10°C for 58 days after initial use.

An analysis of calibrator and QC open vial stability was performed to determine the percentage shift seen between values obtained at day 58 and the expiration date on the vial. A worst case scenario would result in a 2-4% upward shift above the acceptable 10% allowable shift as compared to the initial, expected value. The manufacturer considers that the magnitude of the deviation is not expected to alter the interpretation of the results or to have a clinical impact.

The manufacturer instructs affected users to take the following actions:-

According to the manufacturer, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 30 November 2015

Back