Medical Device Safety Alert: Beckman Coulter IMMAGE Immunochemistry Systems Cuvette Sectors

01 Mar 2017

Medical device manufacturer, Beckman Coulter, has issued a medical device safety alert concerning its IMMAGE Immunochemistry Systems Cuvette Sectors [Part number (P/N): 470706; all Lot].

The manufacturer has determined that IMMAGE Cuvette Sectors (P/N 470706) may be missing one of two positioning pins. Cuvette Sectors missing a positioning pin can become misaligned. If a Cuvette Sector becomes misaligned, under rare circumstances, the instrument mixer paddle may rub against the inside wall of the Cuvette and create a noise. The manufacturer has no evidence of erroneous results being generated due to this issue to date.

The manufacturer is instructing affected users to take the following actions:-

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.