Medical device manufacturer, Beckman Coulter, has issued a medical device safety alert concerning its DxH 500 Hematology Analyser [Reference: B40601; Software versions 1.0.2].
As a result of internal investigations, the manufacturer has become aware that when the instrument temperature (<18°C to >36.5°C) is exceeded, the DxH 500 allows sample processing and result reporting despite the Instructions for Use stating that "The run sample is inaccessible". Results generated in this scenario may be erroneous. Additionally:
If the instrument is operating out of the temperature range and the warnings on the screen are ignored and the results are used, then:
The manufacturer advises affected users to take the following actions:-
This issue will be corrected in a future software release.
According to the local supplier, the affected products are distributed in Hong Kong
If you are in possession of the affected products, please contact your supplier for necessary actions.
Posted on 29 May 2017