Medical device manufacturer, Guangzhou Sunray Medical Apparatus Co., Ltd., has issued a medical device safety alert concerning its Maternal and Fetal Monitor [Model: SRF618X9, STAN S41; Article Number: P1271-05044].
The manufacturer had received several complaints of intermittent inaccurate ultrasound-derived fetal heart rate recordings by the Maternal and Fetal Monitor (Model: SRF618X9, software version 3.5 and 3.8).
According to the manufacturer, the root cause has been identified to be software-related. The inaccurate ultrasound-derived fetal heart rate recordings may lead to none or delayed intervention, resulting in severe injury or death to the fetus.
The manufacturer has corrected this issue in software version 3.9 which will:
The affected users are advised to take the following actions:
According to the local supplier, the affected products are not distributed in Hong Kong.
If you are in possession of the affected products, please contact your supplier for necessary actions.
Posted on 22 June 2018