Medical Device Safety Alert: Abbott Alinity hq Analyzer

14 Feb 2019

Medical device manufacturer, Abbott, has issued a medical device safety alert concerning its Alinity hq Analyzer [List Number: 09P68-01].

The manufacturer has identified that a patient sample result with an "MCHC Out of Range. Check Sample Integrity" flag on the Alinity hq analyzer may indicate that the processed sample contained a clot.

According to the manufacturer, if the analyzer is not halted by the user, the system will continue to process samples. Subsequent results may be impacted and may not be flagged. Patient Complete Blood Count (CBC) results that are generated following the aspiration of a clot may be incorrect (falsely decreased).

The manufacturer is continuing to investigate this issue and future system upgrades are in development. In the meantime, the affected users are advised to take the following actions:

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 14 February 2019