Medical device manufacturer and distributor, Brainlab AG, has issued a medical device safety alert concerning its RT Elements Radiation Treatment Planning Software [Product Reference: Cranial SRS 1.0.0/1.5.0; Spine SRS 1.0.0/1.5.0; Multiple Brain Mets SRS 1.5.0; RT QA 1.0.0/1.5.0].
The manufacturer has determined that dose distribution calculation by Brainlab RT Elements software may potentially be incorrect when using the Pencil Beam algorithm on the GPU (graphics card) for large tissue models. There has been no report to Brainlab by any user site of undesirable dose distribution delivered to a patient due to this issue at the time of writing.
According to the manufacturer, a risk to patient health cannot be excluded if the issue would re-occur with the affected product versions of Brainlab RT Elements Radiation Treatment Planning Software.
For plans calculated using the affected Pencil Beam algorithm on the GPU, the following data can be reviewed in order to determine whether a patient treatment plan (or QA plan) is affected.
The affected users are advised to ensure for all affected applications that the clinically used CT scans do not create tissue models bigger than 512 MB as explained in points 1 and 2 above.
The manufacturer intends the following corrective actions:
According to the local supplier, the affected products are distributed in Hong Kong
If you are in possession of the products, please contact your supplier for necessary actions.
Posted on 29 March 2019