Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Philips Azurion System with software version 1.2

15 Jul 2019

Medical device manufacturer, Philips, has issued a medical device safety alert concerning its Azurion System with software version 1.2.

The manufacturer has identified the following issues:

Issue 1 – System Cold Restart Time May Be Prolonged

Issue 2 – Poor Image Quality

According to the manufacturer, the software version of the affected product is shown on the start-up screen:

For the Issue 1, a prolonged cold restart time may result in a delay of treatment. If the delay occurs during a critical moment of the procedure, there is a potential for harm for the patient.

For the Issue 2, if the quality of the images obtained during the Cine run is inadequate for diagnostic purposes, the study may have to be repeated, requiring additional x-ray radiation of the patient, the administration of additional contrast agent to the patient, and a prolonged procedure time.

The manufacturer has modified the system software to correct the problem. The updated software (release R1.2.1) will be installed in all affected systems (reference FCO72200430). The affected customers are advised to include the Field Safety Notices with the documentation of the system until the manufacture implements the correction in the system.

For the Issue 1, the customers are also advised to take the following action:

According to the local supplier, the affected products are distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 15 July 2019

Back