Important Safety Alerts

Skip to main content

Safety Alerts and Communications

Important Safety Alerts

Important Safety Alerts

Field Safety Notice: Given Imaging Bravo® pH Monitoring System, part number FGS-0312 (5-Pak) or FGS-0313 (Singles)

12 Jan 2012

Medical device manufacturer, Given Imaging, has issued a field safety notice on Bravo® pH Monitoring System, part number FGS-0312 (5-Pak) or FGS-0313 (Singles).

The manufacturer has received reports of failure of the Bravo capsule to attach to the esophagus or failure of the capsule to detach from the placement device. Occurrence of these failures may result in lengthened time for a Bravo procedure, esophageal injury, and/or the need for additional procedures.

Upon investigations, the manufacturer determined that these failures have a combination of root causes including issues related to tolerances in certain manufacturing dimensions. The manufacturer will replace the affected devices. An "Attention" label on how to perform the rescue procedure and an updated User Manual are now being provided with all Bravo devices.

According to the local supplier, the affected product has been distributed in Hong Kong. If you are in possession of the product, please contact your supplier for necessary actions.

Top
Level Double-A conformance, W3C WAI Web Content Accessibility Guidelines 2.1(Thumbnail) Brand Hong Kong(Thumbnail)