The Medicines and Healthcare products Regulatory Agency (MHRA), United Kingdom posted a field safety notice concerning MICRO-PACE 4580, manufactured by Pace Medical, Inc.
Pace Medical has observed in the REF 4580 that in a specific use of the device the defibrillation protection circuitry may not function as well as intended. If the device is used on a patient, with heart wires and the patient's heart is defibrillated directly on the heart, there may be a chance that the device will be damaged from the defibrillation output current.
The manufacturer has designed and validated a new defibrillation protection circuit which will eliminate the potential for this to occur. Furthermore, it advised that:
For details, please refer to the following MHRA website:
http://www.mhra.gov.uk/Safetyinformation/Safetywarningsalertsandrecalls/fieldsafetynotices/CON185136
If you are in possession of the product, please contact your supplier for necessary actions.
Posted on 25 September 2012