Medical Device Safety Alert: Terumo Advanced Perfusion System 1

24 Dec 2012

Medical device manufacturer, Terumo Cardiovascular Systems Corporation, has issued a medical device safety alert concerning Base for Terumo Advanced Perfusion System 1 (Catalogue number: 801763).

Terumo Cardiovascular Systems (CVS) has become aware of a remote possibility that Terumo Advanced Perfusion System 1 could experience a spontaneous loss of system power. Terumo CVS is issuing a safety notice because the User's Manual for the Terumo System 1 does not provide user instruction for responding to such an event.

According to Terumo CVS, five reports of spontaneous power loss for the Terumo System 1 have been received between 2003 and 2012:

The complete loss of system power would result in loss of all patient support functions, including arterial blood flow, myocardial protection, venting and suction capabilities, safety systems, system alarms and system information. Depending on the response of the clinical team and the availability of backup equipment, a prolonged lack of such support could result in death or serious injury.

The manufacturer supplemented that the potential patient injuries range from no injury to varying degrees of neurologic dysfunction, cardiac dysfunction due to inadequate myocardial protection, end organ dysfunction, or death in the case of an extended period of no flow.

In the event a user experiences a spontaneous loss of power without the automatic initiation of the battery backup system, Terumo CVS recommends:

According to the local supplier, the affected products have not been distributed in Hong Kong.

If you are in possession of the products, please contact your supplier for necessary actions.

Posted on 24 December 2012