Important Safety Alerts

Skip to main content

Safety Alerts and Communications

Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Vascular Solutions Guardian II and Guardian II NC Hemostasis Valves

04 Mar 2013

Medical device manufacturer, Vascular Solutions Inc., has issued a medical device safety alert concerning Guardian II and Guardian II NC Hemostasis Valves. The affected model numbers are FH101, FH101-T, FH101-25, FH101-50, FH102, FH102-T, FH102-25 and FH102-502.

There is a low risk that air may be introduced into certain lots of the device, which may lead to an air embolism. Due to the potential harm, the manufacturer is recalling the affected products. The root cause of the problem is related to a slight reduction in the outside diameter (still within validated tolerances) of the wiper seal of the device.

The local supplier confirmed that the affected products have been distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 4 March 2013

Top
Level Double-A conformance, W3C WAI Web Content Accessibility Guidelines 2.1(Thumbnail) Brand Hong Kong(Thumbnail)