Medical device manufacturer, DePuy (a Johnson & Johnson company) has initiated a medical device field safety corrective action concerning Specialist 2 Intramedullary Rod (SP2 IM Rod)(part number966120).
DePuy has identified the potential for the SP2 IM Rod to fail due to fatigue when excess leverage is applied at the tip. There is a deep J shaped groove at the tip of the rod, which allows a sleeve to lock into place when used in revision cases. It is at the top of this groove that fracture can occur. DePuy has received 9 complaints since 2008 regarding the tip breaking with 8 complaints where the tip was left in the patient.
In remote circumstances, the possible clinical implications related to the SP2 IM Rods fracturing with tips left in the patient include:
Due to the potential for the rod to break, leaving fragments in the patient, a field safety notice was sent out to the affected users to provide additional information on how to use the SP2 IM rods to minimize the potential for breakage. In additional, DePuy is currently investigating a material change to the rod to reduce the possibility of the tip fracturing.Changes have/will be made to the surgical techniques to include the guidance below.
DePuy would like to emphasize several technical points regarding the use of the rod that may further reduce the incidence of tip fracture:
According to the local supplier, the affected products were distributed in Hong Kong.
If you are in possession of the affected product, please contact your supplier for necessary actions.
Posted on 6 March 2013