Medical Device Safety Alert: Roche NH3L on cobas c 501/502

13 Mar 2013

Medical device manufacturer, Roche has initiated a medical device field safety corrective action concerning NH3L (catalogue no.: 20766682322) on cobas c 501/502.

The manufacturer is providing a workaround to the affected users if the users have experienced elevated QC or patient recovery over time for the above affected product.

Blank signal may increase due to NH3 formation in R3 as a consequence of NADPH decomposition over the shelf-life of the reagent. This leads to falsely elevated test results over time.

The manufacturer advises users to follow the following procedures if they have the problem of positive drifting recovery of NH3L.

Before use, every new reagent cassette has to be pre-conditioned outside the analyzer for 24 hours at room temperature to allow the NH3 to dissipate from the reagent.

According to the manufacturer, the investigations demonstrated the above procedures can maintain a stable recovery of the assay for 3 days.

According to the local supplier, the affected products were distributed in Hong Kong.

If you are in possession of the affected product, please contact your supplier for necessary actions.

Posted on 13 March 2013