Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Hospira Gemstar Infusion Pump

25 Mar 2013

Medical device manufacturer, Hospira Inc has initiated a field safety corrective action concerning Gemstar infusion pump which four potential risks were identified to be related to the device. The affected list numbers are 13000, 13150 and 13087.

According to Hospira, the affected pumps may have the following problems:

(1) Gemstar lithium battery - low voltage

(2) Gemstar battery leakage

(3) Gemstar pressure sensor calibration drift

(4) Gemstar backward motor movement

Hospira warns the users that a delay in therapy may be resulted. The affected users are advised to:

The local supplier confirmed that the affected products have been distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 25 March 2013

Back