Medical Device Safety Alert: Philips HeartStart MRx Monitor/Defibrillator

19 Apr 2013

Medical device manufacturer, Philips Healthcare, has issued a field safety notice concerning HeartStart MRx Monitor/Defibrillator [Model no.: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6; Serial numbers within the range US00100100 to US00565942]

According to the manufacturer, the Philips HeartStart MRx Monitor/Defibrillator may fail to deliver defibrillation therapy in either Manual Defib or AED mode. If this occurs, the MRx may simultaneously display a "No Shock Delivered" message along with a "Shock Equip Malfunction", INOP and a Red X in the RFU indicator. In addition, the pads ECG waveform may display a non-physiologic flat line rhythm. If the problem occurs, there may be a failure to deliver shock therapy to a patient in need of defibrillation or cardioversion therapy.

The manufacturer will provide software upgrade to fix the problem. However, prior to performing the software upgrade, the manufacturer recommends customer to follow step 1.

If Step 1 does not resolve the problem follow step 2.

According to the manufacturer, the affected products were distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 19 April 2013