Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Medtronic Paradigm Insulin Infusions Sets

13 Jun 2013

Medical device manufacturer, Medtronic Inc., has issued a medical device safety alert concerning the Paradigm insulin infusion sets [Model no.: MMT-317, MMT-318, MMT-324, MMT-325, MMT-312S, MMT-312L, MMT-386, MMT-387, MMT-394, MMT-396, MMT-397, MMT-398, MMT-399, MMT-377, MMT-378, MMT-381, MMT-382, MMT-383, MMT-384, MMT-368, MMT-862, MMT-864, MMT-866, MMT-874, MMT-876, MMT-884, MMT-886, MMT-921, MMT-923, MMT-925, MMT-941, MMT-943, MMT-945, MMT-961, MMT-963, MMT-965, & MMT-975].

The manufacturer has become aware of a potential safety issue that can occur if insulin or other fluids come in contact with the inside of the connector on Medtronic Paradigm infusion sets.

Exposure of the inside of the tubing connector to fluid is most likely to occur if insulin is spilled on the top of insulin reservoir when the reservoir is removed from the transfer guard after filling the reservoir from a vial of insulin. If this occurs, the insulin can temporarily block the vents in the connector that allow the pump to properly prime. If these vents are blocked, this can potentially result in too much or too little insulin being delivered, which may cause hypoglycemia or hyperglycemia, which, in extreme cases, may cause loss of consciousness or death. However, the manufacturer has not been aware of any adverse incidents related to aforementioned potential safety issue.

The potential for temporary blocking of the infusion set vents can be avoided by not allowing liquid to make contact with the inside of the tubing connector and following the recommended reservoir filling procedure with careful attention to the information provided below:

In addition, the manufacturer is also updating the instructions for use to include this information on how to avoid temporary blocking of the infusion set vent membranes due to exposure to fluid.

According to the local supplier, the affected products were distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 13 Jun 2013

Back