Medical Device Safety Alert: GE Healthcare Tec 6 and Tec 6 Plus Vaporizers

13 Aug 2013

Medical device manufacturer, GE Healthcare issued a medical device safety alert concerning all Tec 6 and Tec 6 Plus Vaporizers.

According to the GE Healthcare, the user manual pre-operative check, low pressure leak test at 1% dial setting may not detect the full range of leaks from seal wear degradation in the vaporizers. Moreover, the integrity of the vaporizer seal may be affected with Tec 6 or Tec 6 Plus Vaporizer remanufactured/ manufactured between 12 Aug 2005 and 3 Dec 2012, due to wear degradation.

For any Tec 6 or Tec 6 Plus Vaporizer remanufactured / manufactured between 12 Aug 2005 and 3 Dec 2012 (AETJ29001 to AESR49001), a GE Healthcare Service Representative will schedule a repair to update the vaporizer seal. The leak test failures will get first priority for scheduling repair/replacement

The manufacturer advises that the addendum with the field safety notice distributed to the affected users had the 12% low pressure leak test which will aid in determining the integrity of the seal. The manufacturer also provides the following recommendations to the affected users:

According to the local supplier, the affected products have been distributed to Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.