Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: GE Healthcare Precision 500D, Legacy, RFX, and SFX X-ray imaging systems

02 Oct 2013

The Medicines and Healthcare products Regulatory Agency (MHRA), United Kingdom posted a field safety notice concerning Precision 500D, Legacy, RFX, and SFX X-ray imaging systems, manufactured by GE Healthcare. The affected part numbers are 2258627, 2305472, 2305495, 2404103, 2403791, 2403791-3, 2403791-32, 2403791-40, 46-262751G2, 46-262751G4, 46-262751G5, 46-262751G6, 46-262751G7, and 46-262751G8.

There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This issue, if present, can result in a loss of calibration and offset of the power assist handle and/or, under extreme condition, sticking of the handle in the ON position, and causing the system to strike the patient. No injuries have been reported to date related to this issue.

The manufacturer advises users to inspect power assist handle for either of the two conditions listed below, and immediately contact local GE Healthcare Service Representative if any issue is found.

Furthermore, GE Healthcare will correct all affected systems.

For details, please refer to MHRA website
http://www.mhra.gov.uk/Safetyinformation/Safetywarningsalertsandrecalls/fieldsafetynotices/CON316301

If you are in possession of the product, please contact your supplier for necessary actions.

Posted on 2 October 2013

Back