Important Safety Alerts

Important Safety Alerts

Medical Device Safety Alert: Rudolf Medical GmbH laparoscopic instruments

06 Dec 2013

Medical device manufacturer, Rudolf Medical GmbH, has issued a medical device safety alert concerning its laparoscopic instruments. The item numbers and descriptions of the affected products as follows:-

According to the manufacturer, a customer complaint brought to their attention that during reprocessing, the inner lumen of these laparoscopic inserts cannot be flushed out sufficiently with the recommended rinsing tip. Further investigation showed that all inserts, which cannot be disassembled – as listed above - are affected. Risk of infection exists.

The manufacturer advises users of the followings:

According to the manufacturer, the affected products have been distributed in Hong Kong.

If you are in possession of the affected products, please contact your supplier for necessary actions.

Posted on 06 December 2013

Back