Medical Device Safety Alert: Philips HeartStart MRx Monitor/Defibrillators

26 Feb 2014

Medical device manufacturer, Philips Healthcare, has issued a field safety notice concerning HeartStart MRx Monitor/Defibrillator. The affected model numbers and serial numbers are as follows:

According to the manufacturer, a component of the MRx Processor Board may be susceptible to damage from electrostatic discharge. Damage to these components can disrupt electrocardiography (ECG) and pulse oximetry (SpO2) functionality leading to an inability to perform:

The presence of this condition will be displayed on the ready for use indicator as a solid Red X with periodic audio chirp.

A hardware upgrade will be provided to correct the issue. The manufacturer advises healthcare professionals that the affected product can continue to be used prior to receiving the hardware update. The manufacturer also provides instruction for the appropriate actions to be taken if the affected product displays a solid red 'X'.

According to the local supplier, the affected products were distributed in Hong Kong.

If you are in possession of the affected product, please contact your supplier for necessary actions.

Posted on 26 February 2014