Medical device manufacturer, Philips Healthcare., has issued a medical device safety alert concerning its Incisive CT/CT 3500/CT 5300 System [HKMD No.: 180534].
The manufacturer has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT/CT 3500/CT 5300 system. This component’s structural integrity may become compromised, resulting in potential contact with other gantry components located within the Incisive CT/CT 3500/CT 5300 system during rotation. Other components could be damaged due to the contact with the detached component.
If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs:
If you are in possession of the affected products, please contact your supplier for necessary actions.
Identified on 25 Apr 2024