Safety Alerts

Safety Alerts

Starting from 1 January 2022, active safety alerts related to medical devices listed under MDACS are listed below. For safety alerts before 1 January 2022, please refer to archive.

Medical Device Safety Alert: Philips Healthcare Incisive CT/CT 3500/CT 5300 System

25 Apr 2024

Medical device manufacturer, Philips Healthcare., has issued a medical device safety alert concerning its Incisive CT/CT 3500/CT 5300 System [HKMD No.: 180534].

The manufacturer has identified a hardware issue with a metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT/CT 3500/CT 5300 system. This component’s structural integrity may become compromised, resulting in potential contact with other gantry components located within the Incisive CT/CT 3500/CT 5300 system during rotation. Other components could be damaged due to the contact with the detached component.

If this issue occurs, a loud noise will be emitted and the system will shut down. The cover may become displaced if the following sequence occurs:

If you are in possession of the affected products, please contact your supplier for necessary actions.

Identified on 25 Apr 2024
Posted on 26 Apr 2024

Back