Medical Device Safety Alert: Siemens Healthcare Diagnostics Inc. ADVIA Centaur CA 19-9

05 Aug 2024

The Australian Therapeutic Goods Administration (TGA) has posted a medical device safety alert concerning Siemens Healthcare Diagnostics Inc. ADVIA Centaur CA 19-9 [Lots: 55970535, 55971535, 55972535, 55973535] [HKMD number: 210417].

According to the TGA, there is a positive bias in some samples around a concentration of 37 U/mL with lots 55970535, 55971535, 55972535, 55973535 when compared to the previous lot. A sample bias was observed in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.

For further information, please visit: https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2024-RN-00652-1

If you are in possession of the affected products, please contact your supplier for necessary actions.