Medical Device Safety Alert: Philips Medical Systems Nederland B.V. Spectral CT systems or the Spectral CT on Rails systems
15 Nov 2024
The Australian Therapeutic Goods Administration (TGA) has posted a medical device safety alert concerning Philips Medical Systems Nederland B.V. Spectral CT systems or the Spectral CT on Rails systems [Product code: 728334, 728333 and 728340] [HKMD No.: 220236].
According to the TGA, the manufacturer has identified two problems that could affected the performance of the Philips Spectral CT systems or the Spectral CT on Rails systems.
Problem 1: An interruption during the "get ready for scan" stage of a scan or an interruption during air calibration may occur. If this occurs it may require a system reboot or if this occurs during a scan it could result in a partial set of images and patient may require a rescan.
Problem 2: A synchronisation problem in the Firmware can result in an image quality problem where a dead layer will appear as a ring/band artifact. If this occurs during a scan it could result in a partial set of images and patient may require a rescan.
For further information, please visit: https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2024-RN-01022-1
If you are in possession of the affected products, please contact your supplier for necessary actions.
Posted on 15 Nov 2024